Maintaining a Clean Sample Surface:
Gloves: Use only polyethylene gloves. Other gloves may contain silicones that can contaminate the surface.

Clean: Use only clean utensils (tweezers, etc) when handling samples. Make sure theses have been cleaned to remove hydrocarbon and silicone contaminants and dedicate their use to SIMS samples only. Keep your samples in a “clean” environment. A laminar flow hood or a clean laboratory environment is strongly suggested.

Common Contaminants: Common surface contaminants include:

  Bullet   Hydrocarbon: Pump oil, greasy finger prints, dirty dessicators, dirty solvents.
  Bullet   Silicones: Non-approved gloves, glass-fitting-grease, dirty dessicators, hair, hand lotion.
  Bullet   Salts: Sodium, chlorine, potassium, all have strong signals in SIMS. Improper rinsing or exposure to water that has not been properly purified.
 
Useful Analysis Requires Good Planning:
Controls: Always include control samples with the sample of interest. A control could include a sample of just the underlying substrate and/or a sample of the underlying substrate exposed to solvent used for the surface modification without the actual modifying agent.

Duplicates: It is best, if possible, to send duplicate samples, even if only one is to be analyzed. Sometimes samples are damaged in transport or, on a very rare occasion, can be damaged while loading for analysis.

Include With Samples:

  Bullet   Samples Summary: Include a sheet with a list of the specific samples that are included and what is on them.
  Bullet   Structure: Proper SIMS analysis requires the structure of the surface bound species and knowledge of the underlying substrate. Without this information accurate analysis of the SIMS data is difficult and sometimes not possible.

 

Figure 1: Iontof sample holders. The backmount (left) can accommodate samples ~1.3cm x 1.3cm square. The topmount (right) can accommodate a wide range of samples.
Sample Shipping:

Packaging: We suggest shipping samples in tissue culture polystyrene (TCPS) dishes or polystyrene or polyethylene tubes sealed with parafilm. To prevent the samples from rattling around during transport when using TCPS dishes, we use a small amount of double sided tape to secure the back of the samples to the bottom of the TCPS dish. It is very important to only stick a small corner of the sample to the tape (Figure 2 left panel). If the center of the samples is stuck to the tape or there is tape under the entire sample it is nearly impossible to remove the samples without damaging them. Please try sticking down a “test” sample and be sure it can easily be removed before packaging other samples. We have shipped many samples without incident and have discovered that only taping a small part of the corner is required to secure the sample.

You can also place your samples into a PS or PE tube. Choose a tube that is slightly larger than your sample width. Only put one sample per tube. This allows safe shipping without anything touching the surface of the sample. You can also backfill the tube with nitrogen and seal the cap with parafilm to further preserve your sample (Figure 2 right panel).
Know Your Contact: Be sure that you are shipping your samples to the correct address and contact person. Our shipping address is NOT the same as the usual mailing address. The shipping address will be given out once samples have been approved for analysis. Also, inform your contact that the samples are being shipped to ensure that they will look out for your samples, and inform you if they do not arrive. If you have questions, please contact us.

Sample Physical Requirements:

Sample Size: General SIMS samples need to be flat surfaces. SIMS is very sensitive to topography changes.

-ION-TOF: The ION-TOF instrument can handle a wide variety of sample sizes and shapes. Please contact NESAC/BIO for details.

Figure 2: Left panel: Only a small piece of tape at the sample corner should be sufficient to fix your samples in place for shipping. Right panel: Samples can also be shipped in a PS or PE tube.